A biosafety cabinet protects the person, the sample and the surroundings. A laminar air flow cabinet protects only the work. A fume hood protects the person from chemical fumes. Choose by what you handle.
1. Start With What You Need To Protect
The three cabinets look alike from the front, but they protect different things. Decide first whether you must protect the sample, the person at the bench, the room around them, or all three. Then match the equipment to the work.
| Biosafety cabinet (Class II) | Laminar air flow cabinet | Fume hood | |
|---|---|---|---|
| Main job | Contains biological hazards and keeps the sample clean | Keeps the work surface clean | Draws chemical fumes away from the person |
| Protects the person | Yes, by air drawn in at the front opening | No | Yes, from fumes and vapours |
| Protects the sample | Yes, with HEPA-filtered air | Yes, with HEPA-filtered air | No, because room air is drawn in |
| Protects the surroundings | Yes, because the exhaust air passes a HEPA filter | No | Fumes are carried away, often outside the building through a duct |
| Typical work | Microbiology, cell culture, handling biological samples | Sterile media preparation, assembling sterile parts, non-hazardous tissue culture | Handling solvents, acids and other volatile chemicals |
| Not suitable for | Volatile toxic chemicals, in a standard recirculating cabinet | Infectious, toxic or hazardous material | Sterile work |
2. Biosafety Cabinet
A biosafety cabinet is a ventilated enclosure that uses HEPA-filtered air. In a Class II cabinet, air is drawn in at the front opening to protect the person, filtered air flows down over the work to protect the sample, and the exhaust air passes a HEPA filter to protect the surroundings. It is the usual choice for microbiology, cell culture and the handling of biological samples. Biosafety cabinets are grouped into Classes I, II and III, and Class II is the most common in laboratories.
3. Laminar Air Flow Cabinet
A laminar air flow cabinet, also called a clean bench, passes HEPA-filtered air over the work surface in one steady direction. It keeps the work free from contamination. It does not protect the person, because the air flows towards the front of the cabinet. Use it for sterile preparation of media, for assembling sterile parts, and for other work with non-hazardous material. A laminar air flow cabinet is not a biosafety cabinet, and you should not use it for infectious or toxic material.
4. Fume Hood
A fume hood draws room air into the hood and carries fumes and vapours away from the person working. Most fume hoods exhaust the air outside the building through a duct. A fume hood protects the person from chemical fumes, but it does not keep the work sterile, because it draws in unfiltered room air.
5. How To Choose
- Handling biological material that could harm people? Choose a biosafety cabinet. Your biosafety officer or risk assessment decides the class of cabinet.
- Keeping a sterile job clean, with non-hazardous material? Choose a laminar air flow cabinet.
- Working with chemical fumes or volatile chemicals? Choose a fume hood.
- Both biological and chemical hazards? Ask your safety officer first. A standard cabinet or hood often does not cover both.
6. Planning A New Laboratory
Alsin offers its own biosafety cabinets, laminar air flow cabinets, fume hoods and complete laboratory setups, together with ELETOM79 instruments. Tell us the work your lab will do, the space you have and your power supply, and we will recommend the equipment. Specifications, prices and delivery times are given in your quote. See cabinets, fume hoods and lab setups from Alsin.
Tell Us About Your Application
Share what you need to run and your lab’s power supply. We will recommend a model and send a quote.
Request a quoteChat on WhatsApp